76
Manganese LR
Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent
solution to each cuvette, replace the caps and invert gently
to mix for about 30 seconds.
Add 1 mL of the HI 93748C-0 0.1% PAN indicator
solution to each cuvette, replace the caps and shake
gently.
Add 3 drops of Dispersing Agent (HI 93703-51) to each
cuvette, replace the caps and invert gently to mix for
about 30 seconds.
Place the cuvette with the reacted deionized water (blank)
into the holder and close the lid.
Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for
2 minutes and then press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.