10 mL
Ammonia LR
SPECIFICATIONS
Range 0.00 to 3.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93700A-0 First Reagent 4 drops (6 drops for seawater)
HI 93700B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93700-01 Reagents for 100 tests
HI 93700-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Ammonia LR
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
AMMONIA LOW RANGE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83099 and is the answer not in the manual?

Hanna Instruments HI 83099 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83099
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals