71
Magnesium
SPECIFICATIONS
Range 0 to 150 mg/L
Resolution 5 mg/L
Accuracy ±5 mg/L ±3% of reading at 25 °C
Light Source Tungsten lamp with narrow band interference filter @ 466 nm
Method Adaptation of the Calmagite method
.
REQUIRED REAGENTS
Code Description Quantity
HI 93752A-0 Mg Magnesium Buffer Reagent 1 mL
HI 93752B-0 Mg Magnesium Indicator Reagent 9 mL
REAGENT SETS
HI 937520-01 Reagents for 50 tests
HI 937520-03 Reagents for 150 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the COLORED OR TURBID SAMPLES
procedure on page 17.
• Select the
Magnesium
method using the procedure described in the
Method Selection
section (see page 12).
• Using one 1 mL syringe add exactly 1.00 mL of HI 93752A-0
Mg Buffer reagent to the cuvette and use the pipette to fill the
cuvette up to the 10 mL mark with the HI 93752B-0 Mg
Indicator reagent.
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
MAGNESIUM