Chlorine Dioxide
CHLORINE DIOXIDE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.10 mg/L ±5% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the Chlorophenol Red method. The reaction between chlorine dioxide and
reagents causes a colorless to purple tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93738A-0 Reagent A 1 mL
HI 93738B-0 Dechlorinating Reagent B 1 packet
HI 93738C-0 Reagent C 1 mL
HI 93738D-0 Reagent D 1 mL
REAGENT SETS
HI 93738-01 Reagents for 100 tests
HI 93738-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
• Select the
Chlorine Dioxide
method using the procedure
described in the
Method Selection
section (see page 12).
• Fill two graduated mixing cylinders (# 1 & # 2) up to
the 25 mL mark with the sample.
• Add 0.5 mL of HI 93738A-0 Chlorine Dioxide Reagent to each cylinder (# 1 & # 2), close them and
invert several times to mix.
# 2
25 mL
# 1
25 mL
# 1 & # 2
# 1 # 2