Dissolved Oxygen
SPECIFICATIONS
Range 0.0 to 10.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.4 mg/L ±3% of reading at 25 °C
Typical EMC ± 0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition
,
Azide modified Winkler method. The reaction between dissolved oxygen
and the reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93732A-0 Reagent A 5 drops
HI 93732B-0 Reagent B 5 drops
HI 93732C-0 Reagent C 10 drops
REAGENT SET
HI 93732-01 Reagents for 100 tests
HI 93732-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Dissolved
Oxygen
method using the procedure described in
the
Method Selection
section (see page 12).
Fill one 60 mL glass bottle completely with the unreacted sample.
Replace the cap and ensure that a small part of the sample spills
over.
Remove the cap and add 5 drops of HI 93732A-0 and 5 drops of
HI 93732B-0.
Add more sample, to fill the bottle completely. Replace the cap
again and ensure that a part of the sample spills over. This is to
make sure that no air bubbles have been trapped inside, which
could alter the reading.
Invert several times the bottle. The sample becomes orange-yellow
and a flocculent agent will appear.
DISSOLVED OXYGEN