99
COD MR
Select the
Oxygen Demand, Chemical
MR (COD)
method following one of the
procedures described in the
Method
Selection
section (see page 12).
Place the COD vial adapter in the cuvet
holder and ensure that the adapter is
well fit inside.
Place the blank vial into the holder and push it completely down.
Press the Zero key and the instrument will perform a zero sequence. If the zero sequence was
successfully done, the display will show “-0.0-”. Now the meter is zeroed and ready for measurement.
Remove the blank vial.
Place the sample vial into the holder and push it completely down.
Press Read and the instrument will perform the reading.
The instrument displays concentration in mg/L of oxygen demand on the LCD.
INTERFERENCES
Interference may be caused by:
Chloride (Cl¯) above 2000 mg/L.
Samples with higher chloride concentration should be diluted.