44
Chromium VI LR
CHROMIUM VI LOW RANGE
SPECIFICATIONS
Range 0 to 300 μg/L
Resolution 1 μg/L
Accuracy ±1 μg/L ±4% of reading at 25 °C
Typical EMC ±1 μg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1687-92
,
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93749-0 Powder reagent 1 packet
REAGENT SETS
HI 93749-01 Reagents for 100 tests
HI 93749-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Chromium VI LR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette and add the content of one packet of
HI 93749-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.
10 mL