SPECIFICATIONS
Range 0 to 100 mg/L
Resolution 5 mg/L
Accuracy ±5 mg/L ±3% of reading at 25 °C
Light Source Tungsten lamp with narrow band interference filter @ 466 nm
Method Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension
is measured.
REQUIRED REAGENTS
Code Description Quantity
HI 93751-0 Indicator reagent 1 packet
REAGENT SETS
HI 93751-01 Reagents for 100 tests
HI 93751-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the COLORED OR TURBID SAMPLES procedure on page 17.
• Select the
Sulfate
method using the procedure described
in the
Method Selection
section (see page 12).
• Fill a cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
Sulfate
SULFATE
10 mL