Potassium LR
POTASSIUM LOW RANGE
SPECIFICATIONS
Range 0.0 to 20.0 mg/L
Resolution 0.5 mg/L
Accuracy ±3.0 mg/L ±7% of reading at 25 °C
Typical EMC ±0.5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93750A-0 Potassium Reagent 6 drops
HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the COLORED OR TURBID SAMPLES procedure at page 17.
Select the
Potassium LR
method using the procedure
described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of sample, up to the mark.
Add 6 drops of HI 93750A-0 Potassium Reagent, replace
the cap and swirl the solution.
Place the cuvette into the holder and close the lid.
10 mL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83099 and is the answer not in the manual?