Fluoride
FLUORIDE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition
, SPADNS method. The reaction between fluoride and the liquid reagent
causes a red tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93729-0 SPADNS Reagent 4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Fluoride
method using the procedure described in the
Method Selection
section (see page 12).
Add 2 mL of HI 93729-0 SPADNS Reagent to two cuvettes.
Fill one of the cuvettes with distilled water up to the mark,
replace the cap and invert several times to mix.
Fill the other cuvette with sample up to the mark, replace the
cap and invert several times to mix.
Place the cuvette with the reacted distilled water (# 1) into
the holder and close the lid.
# 1
# 1
# 2
10 mL10 mL
2 mL