EasyManuals Logo

Hanna Instruments HI 83099 User Manual

Hanna Instruments HI 83099
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
55
Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, wait
for two minutes and press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Insert the other cuvette (# 2) with the reacted sample into the
instrument.
Press Read to start reading. The instrument displays the results in mg/L of fluoride.
Note: For wastewater or seawater samples, before performing measurements, distillation is required.
For most accurate results, use two graduated pipettes to deliver exactly 8 mL of distilled water and
8 mL of sample.
Fluoride
# 2
INTERFERENCES
Interferences may be caused by:
Alkalinity (as CaCO
3
) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83099 and is the answer not in the manual?

Hanna Instruments HI 83099 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83099
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals