122
Silver
SILVER
SPECIFICATIONS
Range 0.000 to 1.000 mg/L
Resolution 0.005 mg/L
Accuracy ±0.020 mg/L ±5% of reading at 25 °C
Typical EMC ± 0.001 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm.
Method Adaptation of the PAN method. The reaction between silver and reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93737A-0 Buffer Reagent A 1 mL
HI 93737B-0 Buffer Reagent B 1 mL
HI 93737C-0 Indicator Reagent C 2 mL
HI 93737D-0 Fixing Reagent D 2 mL
HI 93703-51 Dispersing Agent 4-6 drops
REAGENT SETS
HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Silver
method using the procedure described in the
Method Selection
section (see page 12).
Note: for best results perform your tests between 20-24°C.
Fill two graduated beakers with 25 mL of sample.
Add 1.0 mL of HI 93737A-0 Buffer reagent to one beaker
(the blank) and swirl gently to mix.
# 1
# 2
# 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83099 and is the answer not in the manual?

Hanna Instruments HI 83099 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83099
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals