59 Hardness Mg
MAGNESIUM HARDNESS
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.11 mg/L ±5% of reading at 25 °C
Typical EMC ±0.02 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition,
EDTA colorimetric method. The reaction between magnesium and reagents
causes a reddish-violet tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93719A-0 Mg indicator 0.5 mL
HI 93719B-0 Alkali solution 0.5 mL
HI 93719C-0 EDTA solution 1 drop
HI 93719D-0 EGTA solution 1 drop
REAGENT SETS
HI 93719-01 Reagents for 100 tests
HI 93719-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Magnesium Hardness
method using the procedure
described in the
Method Selection
section (see page 12).
Rinse a graduated beaker several times with unreacted sample,
before filling it to the 50 mL mark with the sample.
Add 0.5 mL of HI 93719A-0 Magnesium indicator solution,
then swirl to mix.
Add 0.5 mL of HI 93719B-0 Alkali solution and swirl to mix.
Use this solution to rinse 2 cuvettes.