Hanna Instruments HI 83099 - MEASUREMENT EXECUTION AND INTERFERENCES

Hanna Instruments HI 83099
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74
Manganese HR
Add the content of one packet of HI 93709B-0 Sodium
Periodate reagent. Replace the cap and invert to mix with
gently movements for 2 minutes.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and 30 seconds and press Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of manganese.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of potassium permanganate (KMnO
4
) and
permanganate (MnO
4
¯).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L.