Hanna Instruments HI 83099 - MEASUREMENT EXECUTION AND INTERFERENCES

Hanna Instruments HI 83099
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
Remove the cuvette and add the content of one packet of
HI 93716-0 DPD reagent. Replace the cap and shake gently
for about 20 seconds to dissolve most of the reagent.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press Read. When the timer ends the meter will perform the
reading.
The instrument displays the results in mg/L of bromine.
INTERFERENCES
Interference may be caused by: Chlorine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
2 minutes after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Bromine