32
Press Timer or wait for 5 minutes. Then invert again
the cuvette 10 times to mix (about 15 seconds).
Reinsert the cuvette into the instrument.
Press Read to start the reading. The instrument displays the results in mg/L of Calcium.
INTERFERENCES:
Interferences may be caused by:
Acidity (as CaCO
3
) above 1000 mg/L
Alkalinity (as CaCO
3
) above 1000 mg/L
Magnesium (Mg
2+
) above 400 mg/L
Replace the cap and invert several times to mix.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette.
Using the 1 mL syringe, add exactly 1 mL of the
HI 93752B-0 Calcium Oxalate Reagent. Replace the
cap and invert the cuvette 10 times to mix (about
15 seconds).
Calcium