Total Chlorine
Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette.
Powder reagents procedure
Add 1 packet of HI 93711 DPD reagent. Replace the cap and
shake gently for 20 seconds (or 2 minutes for seawater analysis).
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of total
chlorine.
Liquid reagents procedure
To an empty cuvette add 3 drops of
HI 93701A-T DPD1 indicator, 3 drops of
HI 93701B-T DPD1 buffer and 1 drop of
HI 93701C DPD3 solution. Swirl gently to
mix and immediately add 10 mL of unreacted
sample. Replace the cap and shake gently
again.
Reinsert the cuvette into the instrument.