II. Installation: Confirm MRI, Ambient Conditions, and Chiller
Model
MRI Model
Details: Confirm the MRI model and confirm the gradient coil version (if applicable). Contact Canon
for assistance.
Chiller Model
Instructions: Haskris provides several chiller models for cooling Canon MRIs. Use the MRI and
gradient coil information, and site maximum ambient conditions, to identify the appropriate chiller
model. These instructions include details applicable to all chiller models and are intended to be used
in conjunction with the specification document which contains model-specific information.
New MRI Installations
Haskris
Model
MR
Ambient Air
OPC24
Galan
Orian STD/XGO
-30°F to +120°F
OPC16
Galan
Orian STD*/XGO
-30°F to +105°F
OPC15
Orian STD
Fortian
-30°F to +120°F
OPC5
Elan
-30°F to +120°F
*For Orian with STD gradient only, OPC16 is
compatible with ambient air up to 120°F
MRI and Chiller Models New MRI Installation