Mechanical installation
PAD4002A A4243-1.0 HBM: public 17
5 Mechanical installation
5.1 Important precautions during
installation
S Welding currents must not be allowed to flow over the
electronics. If there is a risk that this might happen,
you must provide a suitable low-ohm connection to
electrically bypass the electronics.
Notice
PAD electronics are precision measuring elements, and
need to be handled carefully. Dropping or knocking them
may cause permanent damage. Tighten the plug during
installation with a torque of max. 4Nm.
Important
Before mounting several weighing modules into an in
stallation with a bus system, take the following into ac
count:
The printed production number (type plate) is required for
setting up data communication. If the type plate can no
longer be seen after installation, the numbers of each
weighing module should be noted beforehand. This en
ables different addresses to be assigned during the initial
start-up.
Alternatively, before connection to the bus system, you
can connect each weighing module individually with a
PC, in order to set different addresses (see ADR com
mand in the Online Help).