AB1401en_SARev.: 04 / 11.2023
38 / 44
Cleaning and care
Greasing the rubber seal of
the centrifuging chamber
Greasing the rubber seal of
the biosafety system
Trunnion greasing
Checking the accessories
Checking the biosafety
system
Inspecting the centrifuging
chamber for damage
Greasing the motor shaft
Swing-out rotors
Aluminium angle rotors
Metal buckets
Lid with bioseal
Inserting
No statement can be made about the resulting degree of sterility.
The lids of the rotors and bucket must be removed before autoclaving.
Autoclaving accelerates the ageing of materials. It may cause changes to
colours. After autoclaving, the rotors and accessories are to be visually
inspected for damage and any damaged parts are to be replaced immedi-
ately.
The sealing ring in question is to be replaced if there are signs of cracking,
embrittlement or wear. For lids with non-replaceable sealing rings, the whole
lid must be replaced.
The sealing rings must be replaced after autoclaving to ensure the tightness
of the biosafety systems.
8.5 Maintenance
Rub the sealing ring lightly with a rubber care product.
Rub the sealing ring lightly with a rubber care product.
1. Remove accessories.
2. Clean the trunnions.
3. Remove any detergent residues with a damp cloth after using deter-
gents.
4. Grease the trunnions and suspension with Hettich Tubenfett 4051.
5. Excess grease in the centrifuging chamber must be removed.
1. The accessories are to be checked for wear and corrosion damage.
2. Check that the rotor is rmly seated.
1. Visually check all parts of the biosafety system for damage.
2. Check the correct installation position of the sealing ring(s) of the
biosafety system.
3. Replace the damaged parts of the biosafety system.
4. Replace the sealing ring in question immediately if there are signs of
cracking, embrittlement or wear. For lids with non-replaceable sealing
rings, the whole lid must be replaced.
Check the centrifuging chamber for damage.
1. Remove accessories.
2. Clean the motor shaft.
3. Remove any detergent residues with a damp cloth after using deter-
gents.

Table of Contents

Other manuals for Hettich UNIVERSAL 320 R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich UNIVERSAL 320 R and is the answer not in the manual?

Hettich UNIVERSAL 320 R Specifications

General IconGeneral
BrandHettich
ModelUNIVERSAL 320 R
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals