AB1401en_SARev.: 04 / 11.2023
7.2 Programming
7.2.1 Write protection for programs
The programs can be protected to prevent unintentional changes.
When the rotor is at a standstill, write protection can be enabled or disabled
as follows:
1. Press and hold the
[SELECT]
button.
🡆
‘SOUND/BELL’
is displayed after 8 seconds.
2.
Press the
[SELECT]
button.
🡆
‘LOCK’
is displayed.
3. Use
[Rotary knob]
to set
‘OFF’
or
‘ON’
.
OFF = The programs are not write-protected
ON = The programs are write-protected
4. Press the
[START/IMPULS]
button.
🡆 The setting is stored.
If ON is set:
‘*** lock ***’
is displayed brieîž…y.
If OFF is set:
‘*** ok ***’
is displayed brieîž…y.
7.2.2
Opening or loading programs
1. Use the
[SELECT]
button to select the
‘PROG RCL’
parameter.
2. Use the
[Rotary knob]
to set the desired program location.
3. Press the
[START/IMPULS]
button.
🡆
‘*** ok ***’
is displayed brieîž…y.
The centrifugation data of the desired program location is dis-
played
4. To check the parameters: Press the
[SELECT]
button several times.
5. To exit the parameter indicator: Press the
[OPEN/STOP]
button or do
not press any button for 8 seconds.
7.2.3
Entering or changing programs
1.
Retrieve program.
2.
If required: Press the
[RCF]
button to toggle between RPM and RCF
indicator (
‘> <’
).
3.
If required: Press the
[SELECT]
button to select the desired parameter
and set it with the
[Rotary knob]
.
The parameters t/min and t/sec must be set to 0 using the
[Rotary
knob]
to set continuous operation. Continuous operation is shown in
the indicator with
‘¥’
.
4.
Use the
[SELECT]
button to select the
‘PROG STO’
parameter.
5.
Use the
[Rotary knob]
to set the desired program location.
6.
Press the
[START/IMPULS]
button.
🡆 Settings are stored in the desired program location.
‘*** ok ***’
is displayed brieîž…y.
The settings are always stored in program location # if the
[START/
IMPULS]
button is pressed without the
‘PROG STO’
parameter
being selected.