AB1401en_SARev.: 04 / 11.2023
Parameter query
4. Press the
[START/IMPULS]
button.
🡆 The setting is stored.
‘*** ok ***’
is displayed brieîž…y.
5. Press the
[STOP/OPEN]
button twice or wait 8 seconds to exit the
menu.
7.5 Machine Menu
7.5.1 Querying system information
The rotor is stationary.
1. Press and hold the
[SELECT]
button for 8 seconds.
🡆
‘SOUND/BELL’
is displayed.
2. Press the
[SELECT]
button repeatedly until
‘FU/CCI – S.’
is displayed.
Program version for the frequency inverter
3. Press the
[SELECT]
button repeatedly until
‘HOURS’
is displayed.
Internal operating hours (the time during which the centrifuge was
switched on)
4.
Turn to the right with the
[Rotary knob]
.
🡆
‘STARTS’
is displayed.
Number of centrifugation runs
5. Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG1’
is displayed.
Internal operating hour of the last rotor change
6. Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG2’
is displayed.
Internal operating hour of the penultimate rotor change
7. Turn to the right with the
[Rotary knob]
.
🡆
‘OPhoursCHG’
is displayed.
Internal operating hour of the last operating hours change
8. Turn to the right with the
[Rotary knob]
.
🡆
‘IMBALCHG’
is displayed.
Internal operating hour of the last imbalance cut-off change
9. Turn to the right with the
[Rotary knob]
.
🡆
‘OffsetCHG’
is displayed.
Internal operating hour of the last offset adjustment
10. Press the STOP/OPEN button to exit the menu.
7.5.2 Querying operating hours
The rotor is stationary.
1. Press and hold the
[SELECT]
button.
🡆
‘SOUND/BELL’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘CONTROL:’
is displayed.
🡆
‘CONTROL:’
and the operating hours are displayed.
3. Press the
[STOP/OPEN]
button to exit the menu.