7.6 Message Reference
116
Select and Query the Measurement Mode Setting
Syntax
Command
:FUNCtion <RV/ RESistance/ VOLTage>
Query
:FUNCtion?
Response <RV/ RESISTANCE/ VOLTAGE>
RV.........................ΩV mode
(Resistance and voltage measurement)
RESISTANCE ....Ω mode (Resistance measurement)
VOLTAGE ..........V mode (Voltage measurement)
Example
Command
:FUNC RV
Selects the ΩV mode.
Query
:FUNC?
Response
RV
ΩV mode has been selected.
Set and Query the Resistance Measurement Range
Syntax
Command
:RESistance:RANGe < 0 to 3.1>
Query
:RESistance:RANGe?
Response <300.00E-3/ 3.0000E+0 (NR3)>
Example
Command
:RES:RANG 120E-3
Selects the most suitable resistance measurement range for
measuring 120 mΩ.
Query
Response
:RES:RANG?
300.00E-3
The current resistance measurement range is 300 mΩ.
Note Changing the resistance measurement range clears stored measurement
data (memory function).
Set and Query the Voltage Measurement Range
Syntax
Command
:VOLTage:RANGe <-20 to 20>
Query
:VOLTage:RANGe?
Response <20.0000E+0 (NR3)>
Example
Command
:VOLT:RANG 15
Selects the voltage measurement range for measuring 15 V.
Query
Response
:VOLT:RANG?
20.0000E+0
The voltage measurement range is fixed at 20 V (single range).