7.6 Message Reference
132
Set and Query EXT I/O Lock
Syntax
Command
:SYSTem:ELOCk <1, 0, ON or OFF>
Query
:SYSTem:ELOCk?
Response <ON or OFF>
ON........EXT I/O control is disabled (preventing inadvertent
operations from electrical noise).
OFF......EXT I/O control is enabled.
Example
Command
:SYST:ELOC ON
Query
:SYST:ELOC?
Response
ON
Note This function affects only command input.
Set Local Control
Syntax
Command
:SYSTem:LOCal
Note Switches from remote control (REMOTE indicator lit) to local control (by
panel keys).
Save and Load Measurement Values
Syntax
Command
:SYSTem:SAVE <1 to 126>
:SYSTem:LOAD <1 to 126>
Note Attempting to load a panel number that has not been saved results in an
execution error.
Up to 126 measurement configurations can be saved and loaded. Refer
to “Panel Save and Load Functions” for details.
Backup Current Measurement Configuration
Syntax
Command
:SYSTem:BACKup
Description
Command
The current measurement configuration (settings) is backed up
so that when power is turned on the next time, the same
configuration is restored.
Note Saved panel and backup settings are stored in the instrument’s EEPROM.
Be aware that the number of times that the EEPROM can be rewritten is
limited (to about a million times).