7.6 Message Reference
130
Execute Self-Calibration
Syntax
Command
:SYSTem:CALibration
Self-Calibration State and Setting
Command
:SYSTem:CALibration:AUTO <1, 0, ON or OFF>
Query
:SYSTem:CALibration:AUTO?
Response <ON or OFF>
ON... AUTO Self-Calibration selected
(executes approximately every 30 minutes)
OFF MANUAL Self-Calibration selected
Example
Command
:SYST:CAL:AUTO ON
Query
Response
:SYST:CAL:AUTO?
ON
Note Even when AUTO is selected, Self-Calibration can be manually performed
at any time by sending the
:SYSTem:CALibration command.
Set and Query Measurement Value Output Upon Triggering
Command
:SYSTem:DATAout <1, 0, ON or OFF>
Query
:SYSTem:DATAout?
Response <ON or OFF>
ON... Measured values are output automatically when a trigger
occurs.
OFF Measured values are not output.
Example
Command
:SYST:DATA OFF
Query
:SYST:DATA?
Response
OFF
Note • This function is convenient when you want to obtain measured values by
applying EXT I/O trigger input. When this function is enabled and a
footswitch is connected to the
TRIG terminal of the EXT I/O connector, a
measured value is sent to the PC automatically each time the footswitch
is pressed, so there is no need to send a command from the PC to
obtain measurement values.
• Refer to "Measurement Value Formats" for format details of returned
measurement values.
• This function is not available when the GP-IB interface is selected.
See Section 4.10 Measurement Value Output Function (Page 64).