6.1 Starting Recording
104
Starting Recording at a Good Time Division (Interval Time)
1
Set the recording start method on the
[SET 4/8, REC 2] screen to [INTER-
VAL].
2
Press the key on the Mea-
surement screen.
The instrument will enter the standby state.
3
Recording will start automatically when a good time division is reached
based on the save interval time.
Example
If the
Start/Stop key is pressed at 11:22:23 with the interval save time set to 5 min-
utes, the instrument will start recording at 11:25:00.
The recording LED flashing
If the save interval is set to 30 seconds or less, recording will start
from the next :00 seconds.