2.2 Preparing to Use the Instrument after Purchase
24
The battery pack is used to power the instrument during power outages and as a backup
power supply. When fully charged, it can provide backup power for approximately 8
hours in the event of a power outage. Note that if a power outage occurs while the bat-
tery pack is not being used, displayed measurement data will be erased. (Data that has
been recorded on the SD memory card and instrument’s internal memory is retained.)
The battery pack is subject to self-discharge. Be sure to charge the battery pack before
initial use. If the battery capacity remains very low after recharging, the useful battery life
is at an end.
See: "Instrument Installation" (p. 7)
Installing (replacing) the Battery Pack
For battery operation, use only the Hioki Model PW9002 Battery
Set. We do not take any responsibility for accidents or damage
related to the use of any other batteries and/or screws.
To avoid the possibility of explosion, do not short circuit,
disassemble or incinerate battery pack. Handle and dispose
of batteries in accordance with local regulations.
To avoid electric shock, turn off the power switch and discon-
nect the cords and cables before replacing (rem
oving) the
battery p
ack.
After replacing the battery pack, replace the case and
screw
s before using the instrument.
Install the battery case onto the PW3360 using the screws sup-
plied with the PW9002 (M3 x 25 mm), keeping the protector
atta
ched to the case. Installing the battery case with the protec-
tor removed or using screws longer than the accompanying
screw
s
may damage the PW3360.
Do not use the screw holes used for installing the protector or the
battery
case for other purposes. Doing so may damage the prod-
uct.
To remove the 9459 Battery Pack, follow Step 4 to 7 in reverse order.
When removing the PW9002 Battery Set from the back of the
instrument and operating the instrument without the battery pack
installed, attach the protector, following Step 4 to 7 in reverse
order. Attach the protector using the four accompanying screws
(M3×6 mm), which secured the protector onto the instrument
when you received the instrument. Securing the protector using
screws longer than the accompanying screws may damage the
instrument.