12.6 Range Configuration and Accuracy by Clamp Sensor
210
12.6 Range Configuration and Accuracy by
Clamp Sensor
The range-configuration table shows the full-scale display value
of each measurement range.
Voltage measurements is indicated as 5 V to 1000 V. If a mea-
surement is below 5 V, it will be zero-suppressed.
Current measurements is indicated as 0.4% to 1
30% f.s. of the
range. If a measurement is below 0.4% f.s., it will be zero-sup-
pressed.
Power measurement is indicated as 0% to 130% f.s. of the
range.
It will be zero-suppressed when the voltage or current is
0.
The range configuration for apparen
t power (S) and reactive
power (Q) is the same, except that the unit is changed to VA and
var, respectively.
When the VT ratio and CT ratio are set, the ranges will be multi-
plied by (VT ratio x CT ratio) (when a power range falls below
1.
000
0 mW or exceeds 9.9999 GW and a current range falls less
than 1 mA, a scaling error occurs and the setting is not accepted).