2.5 Supplying the Power
33
3
2
Chapter 2 Measurement Preparations
2.5 Supplying the Power
Connecting the AC Adapter
Use only the specified Model Z1006 AC Adapter. AC adapter
input voltage range is 100 to 240 VAC (with ±10% stability) at
50/60 Hz. To avoid electrical hazards and damage to the
instrument, do not apply voltage outside of this range.
To avoid damaging the power cord, grasp the plug, not the cord,
when unplugging it from the power outlet.
Make sure the power is turned off before connecting or discon-
necting the AC adapter.
When using a 250 V power cord, for example with a version of
the instrument sold in China,
Australia, or the EU, it may appear
that the power cord cannot be inserted all the way into the AC
adapter. This is normal. As long as the power cord has been
inserted into the AC adapter until it stops, there is no problem
with the connection.
250V