4.2 Changing Measurement Settings
65
3
Chapter 4 Changing Settings
4
Voltage range
The voltage range is fixed to 600 V.
VT ratio (PT ratio)
Set when using a VT (PT) to perform measurement.
Measurement 2 Setting Screen
Selection
Manual 0.01 to 9999.99
Select 1/60/100/200/300/600/700/1000/2000/2500/5000
When taking measurements on the secondary side of a voltage
transformer (VT), you can set the VT ratio in order to convert the
readings to their primary-side equivalents and display the
results. For a VT with a primary-side voltage of 6.6 kV and a sec
-
ondary-side voltage of 110 V, the VT ratio would be 60 (6,600 V /
110 V).
Since the current range is fixed at 600 V, it would be multiplied by
the VT ratio of 60
to yield a voltage range of 36 kV.