1.2 Features
12
1.2 Features
Quick Set function
The Quick Set function simplifies instrument operation by walking users
through a series of steps to configure basic settings, wirings, wiring check
(wiring confirmation), recording settings, and the start of recording in order to
prevent mistakes.
See: Chapter 7, "Quick Set" (p. 109), Measurement Guide (published separately in color)
Wiring Check (wiring confirmation)
When wirings have been set up improperly, a help function displays hints to
help users establish proper wirings.
See: 3.8, "Verifying Correct Wiring (Wiring Check)" (p. 56)
Ability to make measurements even when power is not available
from a wall outlet
The PW9003 Voltage Line Power Adapter (option) can be used to supply
power from measurement lines.
See: "Supplying Power from Measurement Lines (Using the PW9003 Voltage Line Power
Adapter)" (p. 35)
Ability to operate for about 8 hours on battery power
Even when AC power is unavailable, the optional battery pack can be used to
enable about eight hours of measurement.
See: "Installing (replacing) the Battery Pack" (p. 24)
Corresponding to the various power line
The instrument can perform single-phase/2-wire (up to three circuits), single-
phase/3-wire, 3-phase/3-wire (2-power measurement/3-power measurement),
and 3-phase/4-wire measurement. When performing single-phase/3-wire, or
3-phase/3-wire 2-power measurement, the instrument can perform power and
leakage current measurement simultaneously.
See: 4.2, "Changing Measurement Settings" (p. 63)
Broad operating temperature range
The instrument can be used at temperatures ranging from -10°C to 50°C.
However, the operating temperature range is limited to 0°C to 40°C when
operating on battery power and from 0°C to 50°C when using the LAN.
TFT color LCD
The instrument uses an LCD that is easy to see in both dim and bright condi-
tions.