12.2 Basic Specifications
180
Display range, effective measurement range, effective peak range chart
(representative example: 9661 sensor)
Item Range
Display
range
Effective measurement
range
Display
range
Effective peak
Lower limit Lower limit Upper limit Upper limit Range
Voltage
600 V single range
5.00 V 90.00 V 780.00 V 1000.0 V
±1400 Vpeak
Current
(Model 966
1)
5 A range
0.0200 A 0.2500 A 5.5000 A 6.5000 A
±20 Apeak
10 A range
0.040 A 0.500 A 11.000 A 13.000 A
±40 Apeak
50 A range
0.200 A 2.500 A 55.000 A 65.000 A
±200 Apeak
100 A range
0.40 A 5.00 A 110.00 A 130.00 A
±400 Apeak
500 A range
2.00 A 25.00 A 550.00 A 650.00 A
±1000 Apeak
Display specifications
Display update rate
Approx. 0.5 s (excluding SD memory card, during internal
memory access LAN and USB communications) However,
approx. 1.0 s for energy-related data.
Display 320 x 240 dots, 3.5" TFT color LCD
Language Japanese/ English/ Chinese (simple)
Backlight
LED backlight
AUTO OFF (2 minutes)/ON
The POWER LED flashes during auto-of
f
operation.
Conditions of guaranteed accuracy
Conditions of
guaranteed accuracy
Warm-up time of at least 30 minutes, sine wave input, fre-
quency 50/60 Hz
Temperature and
humidity for
guaranteed accuracy
23°C ± 5°C (73 ± 9°F), 80%RH or less
(applies to all specifications unless otherwise noted)
Display range of
guaranteed accuracy
Effective measuring range
Period of guaranteed
accuracy
1 year