EasyManua.ls Logo

Hioki PW3360-20 - Chapter 13 Maintenance and Service; 13.1 Trouble Shooting

Hioki PW3360-20
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12.5 Calculation Formulas
207
Chapter 12 Specifications
12
* Subscript c: measurement channel, k: order being analyzed; r: resistance after FFT; i: reactance
after FFT
Harmonic voltage, current, and power (PW3360-21 only)
Wiring
setting
Item
Single-phase
2 wire
Single-phase
3 wire
Three-phase
3 wire
Three-
phase
4 wire
1P2W 1P3W 1P3W1U 3P3W2M 3P3W3M 3P4W
Voltage
U
ck
[Vrms]
U
1k
U
1k
U
2k
U
1k
U
1k
U
2k
U
1k
U
2k
U
3k
U
ck
=
When using 3P3W3M wiring, the phase voltage is used.
Harmonic voltage content percentage (%): U
ck
=U
ck
/U
c1
×100 (%)
Current
I
ck
[Arms]
I
1k
I
1k
I
2k
I
1k
I
2k
I
3k
I
ck
=
Harmonic current content percentage (%): I
ck
=I
ck
/I
c1
×100(%)
Active power
P
ck
[W]
P
1k
P
1k
P
2k
P
1k
P
2k
=
-U
1kr
×I
2kr
-U
1ki
×I
2ki
P
1k
P
2k
P
1k
P
2k
P
3k
P
ck
=
U
ckr
×I
ckr
+U
cki
×I
cki
P
k
=P
1k
+P
2k
P
k
=P
1k
+P
2k
+P
3k
Harmonic power content percentage (%): P
ck
=P
ck
/IP
c1
I × 100(%)
Although P
1k
and P
2k
values for 3P3W2M wiring are used in internal calculations, they are
not displayed.
Reactive
power
Q
ck
[var]
Q
1k
Q
1k
Q
2k
Q
1k
Q
2k
=
-U
1kr
×I
2ki
+U
1ki
×I
2kr
Q
1k
Q
2k
Q
1k
Q
2k
Q
3k
Q
ck
=
U
ckr
×I
cki
-U
cki
×I
ckr
Q
k
=Q
1k
+Q
2k
Q
k
=Q
1k
+Q
2k
+Q
3k
Although Q
ck
values are used in internal calculations, they are not displayed.
U
ckr
2
U
cki
2
+
I
ckr
2
I
cki
2
+

Table of Contents