3.7 Setting the Current Range
55
3
Chapter 3 Connecting to Lines to be Measured
3
3
Check the current value and set
the current range.
When measuring multiple single-phase/
2-wire (1P2W) circuits or when “current
only” is selected, you can select other
channels in the same way and set the
range.
Selecting an appropriate range
Select an appropriate range based in-
formation such as the load rating, oper-
ating conditions, and breaker rating. If
the rang
e
is too low, you will experience
an over-range event during measure-
ment, making accurate measurement impossible. Conversely, if the range is too
high, the error
will be too large, making accurate measurement impossible. Set the
current range based on the maximum load current that is anticipated during the
measurement interval.
9660 5A, 10A, 50A, 100A
9661 5A, 10A, 50A, 100A, 500A
CT9667-500A 50A, 100A, 500A
CT9667-5kA 500A, 1kA, 5kA
9669 100A, 200A, 1kA
9694
500mA, 1A, 5A, 10A, 50A
9695-02
9695-03 5A, 10A, 50A, 100A
9657-10
50mA, 100mA, 500mA, 1A, 5A
9675