EasyManua.ls Logo

Hologic DISCOVERY A - Page 194

Hologic DISCOVERY A
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Discovery QDR Series Technical Manual
5-28
Install the ASY-00271 Detector Assembly
1. Install the ASY-00271 Detector Assembly (see Figure 5-13) in place of the 010-
1653 using the screws removed in Step 3.
2. Connect the 180-0292 laser cable.
3. Re-route the 180-0419 ribbon cable and install it on P1 of the ASY-00271.
4. Install the 370-0056 clamp-on ferrite core over the 180-0419 ribbon cable.
5. Turn on the Discovery power, boot the computer and login as Service.
6. Perform “X-Ray Beam Alignment ” on page 4-4.
7. Perform “A/D Gain Control Adjustment” on page 4-26.
8. Perform “Laser Positioning Offset Adjustment” on page 4-26.
9. Replace any covers that were removed and perform “Detector Flattening ” on
page 4-29.
10. Perform the Field Service Calibration procedure.
Figure 5-13. Detector Assembly ASY-00271
11. Perform a System Backup.
12. Perform a System Recover. Verify that the System Backup recovers without error.
13. Verify proper system operation.
ASY-00271
370-0056
180-0419
180-0292
laser cable

Table of Contents

Related product manuals