EasyManua.ls Logo

Hologic DISCOVERY C - Page 163

Hologic DISCOVERY C
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Discovery QDR Series Technical Manual
4-41
Slot=7, 2329 Factory
Slot=8, 1919 Factory
Slot=9, 1704 Factory
Slot=10, 1324 Factory
Slot=11, 1109 Factory
Slot=12, 649 Factory
Slot=13, 349 Factory
SlotOff=0, 3909 Factory Number and Position “Off Aperture”
SlotOff=1, 3909 Factory
SlotOff=2, 3281 Factory
SlotOff=3, 3281 Factory
SlotOff=4, 2714 Factory
SlotOff=5, 2714 Factory
SlotOff=6, 2124 Factory
SlotOff=7, 2124 Factory
SlotOff=8, 1514 Factory
SlotOff=9, 1514 Factory
SlotOff=10, 879 Factory
SlotOff=11, 879 Factory
SlotOff=12, 174 Factory
SlotOff=13, 174 Factory
[Attenuator]
Tolerance=15 Factory +/- Tolerance for Attenuator Blocks
Block=0,4038 Factory Attenuator Block Number and Distance
Block=1,3809 Factory
Block=2,3559 Factory
Block=3,3309 Factory
Block=4,3059 Factory
Block=5,2809 Factory
Block=6,2559 Factory
Block=7,2309 Factory
Block=8,2059 Factory
Block=9,1809 Factory
Block=10,1559 Factory
Block=11,1309 Factory
Block=12,1059 Factory