Discovery QDR Series Technical Manual
4-44
QC_HILO_FACT_MAX =
200.0
Factory Long-term drift check warning message limit.
QC_HILO_FACT_MIN =
100.0
Factory Long-term drift check error message limit.
UPDATED = 06/20/94
12:00:00
Automatic Date and time the ARRC.TXT file was last updated.
UPDATED_N = Automatic Date Last Updated - Normal
UPDATED_NL = Automatic Date Last Updated - Lateral
ARRC.TXT Variables Set by… Descriptions
QDR_SERIAL_NB =
80015
Factory Serial number of the Discovery.
ARRAY_PHANTOM_NB
= 1922
Fact. or F.S. Serial number of the phantom shipped with the
instrument.
ARRC_SEQUENCE_NB =
001
Automatic Sequential number of the last ARRC.TXT file.
ACFT = 1.021550 Fact. or F.S. Area Correction Factor for Turbo scans.
BCFT = 1.005450 Fact. or F.S. Bone Correction Factor for Turbo scans.
SFFT = 1.083600 1.095000 Factory Spine Fan Factors Turbo mode correction factors.
HFFT = 1.042700 1.102800
1.042700 1.102800
1.042700 1.102800
1.042700 1.102800
Factory Hip Fan Factors Turbo mode.
Multiplication factors for BCF & ACF in turbo scan
modes. These values are the same on all Discoverys.
VERSION_NB = 005 3/27/
95
Fan Factor Version # (same for all)
WBINTACF = 1.000 Whole Body Fan Factor Area
WBINTBCF = 1.131 Whole Body Fan Factor BMC
AWBAREA = 0.975 Whole Body Fan Factor Area
AWBBMC = 1.020 Whole Body Fan Factor BMC
AWBLEAN = 0.962 Body Composition Fan Factor lean
AWBWT = 1.01 Body Composition Total Mass
FOREAREA = 1.0 Forearm Fan Factor Area
FOREBMC = 1.0 Forearm Fan Factor BMC
HWVERSION = 14 System Hardware Descriptor
PCAL_STEP_NB = 5 PCAL Step Phantom