Discovery QDR Series Technical Manual
3-11
Figure 3-8. Shipping Bracket Locations (A and SL)
7. Disconnect the ribbon cable from the Arm R Motor Controller board.
8. Remove the Arm Y bearing blocks at the rear of the arm.
9. Remove the bracket that connects the C-arm to the Arm Y belt.
10. Lift the C-arm carriage up and out of the scanner unit.
11. Tape, or tie wrap, the bearing blocks to prevent them from sliding off the end of
the rail.
Figure 3-9. Table Locking “L” Bracket Location
C-Arm Shipping Brackets