5. PC CONFIGURATION
5-25
Error message Error Classification Corrective action
Firmware file does not exist in Error during firmware extraction Ensure that ICOMp.exe and the firmware
files (REC.Z for main firmware and
CBREC.Z for communication board
firmware) are in the location indicated.
Folder contains invalid firmware files Error during firmware extraction Place the correct type and uncorrupted
firmware files in the folder and retry the
operation.
Hardware Error. The Upgrade
Firmware will be aborted
Firmware upgrade error A hardware error has been encountered.
Contact us for assistance.
Invalid entry Error message when values of
parameters are not within valid
range
The value of the parameter is outside its
specified range. Enter a value that is
within the range for the parameter.
Invalid path or file name Error during firmware extraction Check if a proper path is provided to the
firmware file. To avoid typographical
errors use the browse button to locate
the folder containing the firmware files.
Invalid recorder address Invalid recorder address is typed Provide a recorder address which is
within the 0-99 range.
No relay card present in the device! Communication error Insert a relay card into the recorder and
retry the operation.
No Response received from Device Communication error Check connection between PC
Configurator and recorder.
Check that correct recorder type is
selected.
Printer not available Error on non-availability of printer Check if the printer is connected and
configured properly before issuing a print
command.
Protocol Error Communication error Reset the power of recorder and retry the
operation.
Protocol Error. The Upgrade
Firmware will be aborted
Firmware upgrade error A wrong protocol was used for firmware
upgrade. Re-install the PC Configurator
and try again.
Ram Write Error. The Upgrade
Firmware will be aborted
Firmware upgrade error A wrong protocol was used for firmware
upgrade. Re-install the PC Configurator
and try again.
Tag not found Open configuration Provide a tag name from the displayed
list of tag names during an open
operation
The disk is not formatted. Error during an export operation Format the floppy disk and retry the
operation.
The second destination line must be
superior at the first.
Error during an Copy Line
operation
“To” value must be greater than
“Destination Line n” value.
The selected file is not valid for
Upgrade Firmware. The Upgrade
Firmware will be aborted
Firmware upgrade error Chose the correct file for firmware
upgrade. (001XX.REC for main firmware
upgrade and 100XX.REC for
communication firmware upgrade)