Response time (T
90
): 2 seconds
Accuracy (Isobutylene): 10 to 2000 ppm: ±5% at calibration point.
PID Detector: Easy access to lamp and sensor for cleaning and replacement
Calibration: Two-point field calibration of zero and standard reference gases
Patented Reflex PID™ technology
Inlet Probe: Flexible 5" tubing
Keypad: 1 operation key and 2 programming keys; 1 flashlight switch
Direct Readout: Instantaneous, peak value, and battery voltage
Datalogging: 260,000 points with time stamp, serial number, user ID, site ID, etc.
Intrinsic Safety: ATEX
2460 II 2G Ex ia IIC T4 Gb
Ex ia IIB T4 Gb
Sira 17ATEX2082X
IECEx CSA 10.0005X Ex ia IIC T4 Gb
Ex ia IIB T4 Gb
EX ia IIC T4 Gb (with bat pack 059-3051-
000)
EX ia IIB T4 Gb (with bat pack 059-3052-
000)
CSAE 21UKEX2051X
EM Interference: Highly resistant to EMI/RFI.
Alarm Setting: Separate alarm limit settings for Low Alarm and High Alarm
Alarm:
Buzzer 95dB at 30cm and flashing red LEDs to indicate exceeded preset limits, low battery
voltage, or sensor failure
Alarm Type: Latching or automatic reset
Real-time Clock: Automatic date and time
Communication:
Download instrument setup from PC via RS-232 with Travel Charger, or via USB with optional
Charger Stand
Upload data to PC and download instrument setup from PC via USB on charging station.
Enhanced datalogging capability through BLE module and mobile app
Sampling Pump: Internally integrated. Flow rate: 450 to 550 cc/min.
Temperature: -20º C to 50º C (-4º to 122º F)
Humidity: 0% to 95% relative humidity (non-condensing)
Housing (including rubber
boot):
Polycarbonate, splashproof and dustproof
Battery can be changed without removing rubber boot.
Handheld PID Monitors 114 User Manual