The instrument begins charging automatically. (If the optional Charging cradle is
used, the “Primary” LED blinks green to indicate charging.) During charging, the
diagonal lines in the battery icon on the instrument’s display are animated and you
see the message “Charging...”
Note: If the Li-ion battery has been discharged below a certain threshold, the “Charging...”
message does not display immediately. The charging LED blinks to indicate that it is charging,
and after it has been charging for a while, the “Charging...” message appears.
When the instrument’s battery is fully charged, the battery icon is no longer animated and
shows a full battery. The message “Fully charged!” is shown. (If the Charging cradle or Travel
Charger is used, its LED glows continuously green.)
Note: If you see the “Battery Charging Error” icon (a battery outline with an
exclamation mark inside), check that the instrument or rechargeable battery has
been properly set into the Travel Charger (or Charging cradle). If you still receive the
message, check the Troubleshooting section of this guide.
Note: If the instrument or battery has been charging for more than 10 hours and you see the
“Battery Charging Error” icon and a message that says, “Charging Too Long,” this indicates that
the battery is not reaching a full charge. Try changing the battery and make sure the contacts
on the instrument are meeting the Travel Charger’s (or Charging cradle’s) contacts. If the
message is still shown, consult your distributor or Honeywell Technical Services.
WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge and replace
batteries only in areas known to be non-hazardous. Remove and replace batteries
only in areas known to be non-hazardous.
Notes: A rechargeable intrinsically safe Li-ion battery (059-3051-000) for use with
MiniRAE3000+ / ppbRAE3000+ / UltraRAE3000+, a rechargeable Lithium-ion battery (059-
3053-000) (NOT intrinsically safe) for MiniRAE Lite+ only, and an Alkaline battery pack (059-
3052-000) can be charged by placing it directly in the charging port on the back of the
Charging cradle. It can be charged at the same time as the instrument. Press the battery in
place, sliding it slightly toward the front of the Charging cradle. This locks it in the Charging
cradle. To release the battery, slide it forward again and tilt it up.
Note: An Alkaline Battery Adapter (part number 059-3052-000), which uses four AA alkaline
batteries (Duracell MN1500), may be substituted for the Li-Ion battery.