45
SPM Flex Gas Detector
Security
Follow local and site procedures when working with the SPM Flex gas detector�
If needed, verify that the associated control panel is inhibited in order to
prevent false alarms� The following procedures must be followed carefully and
performed only by suitably trained personnel�
When enabled, the detector has five security levels, summarized in this table�
Passcodes are entered at the login menu beginning with the highest level first
(e�g�, set the 4 Admin passcode, then the 3 Power User passcode, then the 2
Advanced Maintenance passcode, etc�)� Store the Admin passcode in a secure
location; if it is forgotten, the detector’s security can be reset only by a Honeywell
Analytics service technician�
Access Rights
Security Levels
0 Not Logged In
1 Routine Maintneance
2 Advanced Maintneance
3 Power User
4 Admin
5 Factory Service
Log in/log out
l l l l l l
Reset alarms and faults
l l l l l l
Review mode
l l l l l l
Limited maintenance access
l l l l l
Change Chemcassette (to same type)
l l l l l
Testing
l l l l
Maintenance
l l l l
Limited setup
l l l l
Full setup
l l l
Change Chemcassette (to any type)
l l l
Security
l l
Program update
l l
Factory service menu (Honeywell personnel only)
l