Installation Model
8904A
HP-lB
Address Selection
The Multifunction Synthesizer is both
a
listener and
a
talker.
Any
address
from
00
to
30
(decimal)
can
be
used,
however, address 21 should not normally be used since
it
is
the address normally
reserved for the internal controller.
HP-IB
addresses greater than
30
are invalid and
will
not
be
allowed
to
be stored in the instrument. The instrument
will
normally be delivered with the
HP-IB
address set
to
26.
The
HP-II3
address
can
only
be
changed while the instrument
is
on,
and by
using
direct front panel
entry. To read
or
change the
HP-IB
address, key-in
SHIFT
ADRS.
The front panel
will
display the
current address and allow the entry
of
a
new address. Enter
a
new HP-IB address by keying-in the
new address and pushing the
ENTER
key.
interconnection
Interconnection data for the Hewlett-Packard Interface
Bus
is
provided
in
Figure 2-1.
Mating Connectors
Coaxial Connectors.
Coaxial
mating connectors used with the Multifunction Synthesizer should
be
50-ohm
BNC
male connectors that are compatible with those specified in
US
MIL-C-39012.
Interface Connector.
HP-IB mating connector is shown in Figure 2-1. Note that the
two
securing
screws are metric.
DIGITAL
PORT
Connector.
This is a 15 circuit, 15 pin, type
"D"
subminiature connector.
Corresponding mating connector is HP part 1251-0221
or
TRW part DAM15P. (The Hop Ram
functions for the port
only
operate on instruments equipped with Option
003
-
Hop
Ram).
Operating Environment
The operating environment should be within the following limitations:
Temperature
............................................................................
.O"
C
to
+
50"
C
Humidity
.......................................................................
<
95% relative
at
40"
C
Altitude..
...................................................................
.<
4570
meters
(15,000
feet)
Bench Operation
The instrument cabinet has plastic feet and
a
foldaway
tilt
stand for convenience in bench operation.
(The plastic feet are shaped
to
ensure self-alignment
of
instruments when they
are
stacked.) The tilt
stand raises the front of the Multifunction Synthesizer for easier viewing
of
the front panel.
Rack
Mounting
2-4
Rack mounting information is provided with the rack mounting kits.
If
a
kit
was not ordered with
the Multifunction Synthesizer as an option,
it
may be ordered through the nearest Hewlett-Packard
office.
rev.
OIJULSI
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com