Getting acquainted
Safety considerations
1-9
m Place the paddles in their holders and depress both Shock buttons.
or
m Leave the unit charged for 60 seconds and it will automatically disarm.
Do not leave the instrument turned on when it is not in use and it is not plugged
into AC power.
Do not discharge the defibrillator with the paddles shorted together. To do so can
cause burning and pitting of the metal paddle contacts.
Disconnect any other medical electronic equipment from the patient during
defibrillation discharge unless labelled as defibrillator protected ( ) ( ).
Avoid open paddle discharges. Dangerous high voltage exists on
the paddles when the defibrillator is discharged. Contact with
this high voltage could cause death or serious injury.
Avoid touching any metal surfaces on the instrument during
shock.
Avoid connecting the patient to several devices at once, because
leakage current limits can be exceeded.
Never touch the bed, the patient, or any equipment connected to
the patient during defibrillation.
Keep the CodeMaster XL+ and the immediate area clean and
dry at all times to avoid creating potentially dangerous electrical
paths.
Never open the instrument case. Dangerous high voltages can be
exposed. Only qualified service personnel can service the
instrument.
Do not use the defibrillator in a flammable or oxygen-rich
atmosphere. This will cause an explosion hazard.
CAUTION
WARNING