Maintaining the Defibrillator
Cleaning Exterior Surfaces
8-5
8 Fully recharge the battery by plugging the instrument into AC power for 18
hours. Verify that the and lights are on.
9 The instrument is now ready to be returned to service.
The battery can be permanently damaged if left uncharged for a prolonged period.
Replacing the Battery
To install a new battery, perform the following steps:
1 Remove AC power from the instrument.
2 Place the instrument upside down on a workbench.
3 Turn the two battery compartment quarter-turn screws counter-clockwise.
4 Lift out the battery door.
5 Disconnect the battery connector and remove the battery.
6 Connect the new battery and slide it into the compartment.
7 Replace the battery door and secure it by turning the two retaining screws a
quarter-turn clockwise.
8 Perform the Battery Capacity Check above before placing instrument in service.
8Cleaning Exterior Surfaces
The CodeMaster XL+ and its accessories are chemically resistant to common
hospital cleaning solutions and non-caustic detergents. The following list includes
some approved cleaning solutions.
o 90% Isopropyl alcohol (except adaptors and patient cable)
CAUTION