Monitoring
Printing the Event Summary Record
3-8
HR alarms are automatically turned off when you press .
3Printing the Event Summary Record
During defibrillator usage, the monitor stores up to 28 ECG strips of critical
information called events. Events include all shocks, heart rate alarms, SpO
2
alarms
and mark events. Each event record includes date of event, heart rate, ECG source,
and size setting as shown in Table 3-6, “Event Summary Record Information,” on
page 3-8. The time annotated on the ECG strip is within 8 seconds of the recorded
event. The message "ES" is printed at the top of the ECG strip when you print the
Event Summary record.
To print the Event Summary on the recorder, press . The recorder must
not be printing to print an Event Summary with this key. After printing an event,
you must wait 10 seconds before printing another event.
To stop printing the Event Summary, press or .
To review the Event Summary later, turn the unit on and press .
Table 3-6 Event Summary Record Information
Event Event Summary Description
Shock Shock#, Delivered energy, peak current, and patient impedance.
Heart Rate
Alarms viola-
tion
Heart Rate alarm limits.
SpO
2
Alarms
violation
SpO
2
alarm limits.
Mark Marker symbol ( ) annotates strip at point was pressed.