Troubleshooting
Operational Checks
7-8
Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
11 Grasp the paddle handles, and without removing the paddles from their holders,
press both Shock buttons simultaneously. A brief automatic recorder run prints
the delivered energy test report.
12 Press to place the instrument in Sync mode.
13 Verify that the messages SYNC and USE LEADS appear on the display.
14 Press once to select Lead I, and verify that the message USE
LEADS no longer appears. The message LEADS OFF should now appear, indi-
cating that one or more leads are not connected.
15 Turn the Energy Select control to Off (Standby).
The defibrillator is ready for use if it passes the above checklist.
A more extensive test of defibrillator/monitor functionality can be performed using
the diagnostic service mode described in the M1722/M1723 CodeMaster XL Series
Defibrillator/Monitor Service Manual.
7Operational Checks
These checks are intended to briefly verify the proper operation of the CodeMaster
XL+ defibrillator/monitor. Regularly perform a test routine incorporating the
following checks along with visual inspection of all cables, paddles, and controls.
Every Shift
Perform the following checks every shift.
WARNING