EasyManua.ls Logo

HP CodeMaster XL+ M1722A - Every Week; Every Three Months; Every Six Months

HP CodeMaster XL+ M1722A
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
Operational Checks
7-12
Every Week
Perform the following checks on the internal paddle set every week.
o Check for excessive residue from sterilization on the paddle set and clean as
needed. Oxidation can be an indication the paddle set is old and must be
replaced.
o Check for pitting or discoloration on the electrode surfaces. Polish or replace as
required.
o Ensure that the cable, connector, and electrodes have no cracks in the
insulation.
Every Three Months
Have the cable set tested for electrical continuity every three months.
Every Six Months
See the M1722/M1723 CodeMaster XL Series Service Manual for extensive
electrical, operational and safety tests to be performed by a qualified Biomedical
Equipment Technician (BMET) or equivalent service technician every three to six
months.