Getting acquainted
AC and DC (Battery) Operation
1-10
Do not rely entirely upon heart rate alarms. Rate meters on
patients can continue to count the pacemaker rates during
cardiac arrest or some arrhythmias. Keep pacemaker patients
under close observation.
Avoid moving a charged defibrillator. If the unit is dropped, it
may discharge.
1AC and DC (Battery) Operation
The CodeMaster XL+ defibrillator can be operated on AC line or DC battery power.
The following is a list of AC and battery operating instructions.
The battery will charge when the instrument is connected to AC power even if the
Energy Select switch is in the Off (Standby) position.
Battery charging is indicated by the light being on.
A fully depleted battery will recharge to 90% of full capacity in two hours and
100% capacity in 18 hours. To preserve battery integrity, the battery must be fully
recharged each time the battery is depleted.
A new battery or one that has been stored for an extended period requires 24 hours
of charging before use.
When the unit is not in use, connect it to AC power with the monitor turned off.
This is to maintain a full battery charge and to prolong battery life.
To operate on internal battery power only, disconnect the power cord from the AC
outlet.
A fully charged battery will nominally provide fifty 360 joule charge-shock cycles
or 2.5 hours of continuous monitoring (15C-40C).