Maintaining the Defibrillator
Supplies
8-8
Ethylene Oxide Sterilization
8Supplies
The Hewlett-Packard warranty is only assured if you use Hewlett-Packard approved
accessories and replacement parts.
Battery
Patient Cables
Temperature
54ºC ± 1ºC (130ºF ± 2ºF)
Relative Humidity
60% ± 20%
EtO Concentration
600 mg/L ± 30 mg/L 12/88 EtO
Exposure Time
1 hour 45 minutes
Aeration Time
18 hours
M1758A
Battery assembly
M1731A
Patient cable 6-pin/3-wire AHA
M1732A
Patient cable 5-wire/6-pin AHA
M1733A
Patient cable 8-pin/3-wire AHA
M1734A
Patient cable 5-wire/8-pin AHA
M1735A
Patient cable 8-pin/3-wire IEC
M1736A
Patient cable 5-wire/8-pin IEC
M1500A
Trunk cable 3-wire/12-pin - AHA
M1605A
Lead set 3-wire - AHA
M1510A
Trunk cable 3-wire/12-pin - IEC
M1615A
Lead set 3-wire - IEC
M1520A
Trunk cable 5-wire/12-pin - AHA
M1625A
Lead set 5-wire - AHA